79 resultados para aberrations, peripheral refraction, ocular aberrations, peripheral aberrations, Hartmann-Shack aberrometry, orthokeratology, intraocular lens implantation, corneal refractive surgery

em QUB Research Portal - Research Directory and Institutional Repository for Queen's University Belfast


Relevância:

100.00% 100.00%

Publicador:

Resumo:

Purpose
To evaluate the impact of the position of an asymmetric multifocal near segment on visual quality.

Setting
Cathedral Eye Clinic, Belfast, United Kingdom.

Design
Retrospective comparative case series.

Methods
Data from consecutive patients who had bilateral implantation of the Lentis Mplus LS-312 multifocal intraocular lens were divided into 2 groups. One group received inferonasal near-segment placement and the other, superotemporal near-segment placement. A +3.00 diopter (D) reading addition (add) was used in all eyes. The main outcome measures included uncorrected distance visual acuity (UDVA), uncorrected near visual acuity (UNVA), contrast sensitivity, and quality of vision. Follow-up was 3 months.

Results
Patients ranged in age from 43 to 76 years. The inferonasal group comprised 80 eyes (40 patients) and the superotemporal group, 76 eyes (38 patients). The mean 3-month spherical equivalent was −0.11 D ± 0.49 (SD) in the inferonasal group and −0.18 ± 0.46 D in the superotemporal group. The mean postoperative UDVA was 0.14 ± 0.10 logMAR and 0.18 ± 0.15 logMAR, respectively. The mean monocular UNVA was 0.21 ± 0.14 logRAD and 0.24 ± 0.13 logRAD, respectively. No significant differences were observed in the higher-order aberrations, total Strehl ratio (point-spread function), or modulation transfer function between the groups. Dysphotopic symptoms measured with a validated quality-of-vision questionnaire were not significantly different between groups.

Conclusion
Positioning of the near add did not significantly affect objective or subjective visual function parameters.

Relevância:

100.00% 100.00%

Publicador:

Resumo:

PURPOSE: To determine whether optical aberrations caused by cataract can be detected and quantified objectively using a newly described focus detection system (FDS). SETTING: The Wilmer Opthalmological Institute, The Johns Hopkins University School of Medicine, Baltimore, Maryland, USA. METHODS: The FDS uses a bull's eye photodetector to measure the double-pass blur produced from a point source of light. To determine the range and level of focus, signals are measured with a series of trial lenses in the light path selected to span the point of best focus to generate focus curves. The best corrected visual acuity (BCVA), refractive error, lens photograph grades, and FDS signals were obtained in 18 patients scheduled to have cataract surgery. The tests were repeated 6 weeks after surgery. RESULTS: The mean FDS outcome measures improved after cataract surgery, with increased peak height (P=.001) and decreased peak width (P=.001). Improvement in signal strength (integral of signal within +/-1.5 diopters of the point of best focus) strongly correlated with improvement in peak height (R(2)=.88, P<.0001) and photographic cataract grade (R(2)=.72, P<.0001). The mean BCVA improved from 20/50 to 20/26 (P<.0001). The improvement in BCVA correlated more closely with FDS signal strength (R(2)=.44, P=.001) than with cataract grade (R(2)=.25, P=.06). CONCLUSIONS: Improvement in FDS outcome measures correlated with cataract severity and improvement in visual acuity. This objective approach may be useful in long-term studies of cataract progression.

Relevância:

100.00% 100.00%

Publicador:

Resumo:

Aims - To evaluate the efficacy of amniotic membrane transplantation (AMT) for ocular surface reconstruction. Methods - 10 consecutive patients who underwent AMT were included. The indications were: group A, cases with persistent epithelial defect after corneal abscess (n = 1), radiation (n = 1), or chemical burn (n = 3); group B, cases with epithelial defect and severe stromal thinning and impending or recent perforation, due to chemical burn (two patients, three eyes) or corneal abscess (n = 2); group C, to promote corneal epithelium healing and prevent scarring after symblepharon surgery with extensive corneo-conjunctival adhesion (n = 1). Under sterile conditions amniotic membrane was prepared from a fresh placenta of a seronegative pregnant woman and stored at -70°C. This technique involved the use of amniotic membrane to cover the entire cornea and perilimbal area in groups A and B, and the epithelial defect only in group C. Results - The cornea healed satisfactorily in four of five patients in group A, but the epithelial defect recurred in one of these patients. After AMT three patients underwent limbal transplantation and one penetrating keratoplasty and cataract extraction. In group B amniotic membrane transplantation was not helpful, and all cases underwent an urgent tectonic corneal graft. Surgery successfully released the symblepharon, promoted epithelialisation and prevented adhesions in the case of group C. Conclusion - AMT was effective to promote corneal healing in patients with persistent epithelial defect, and appeared to be helpful after surgery to release corneo-conjunctival adhesion. Most surgery for further surface rehabilitation. Amniotic membrane used as a patch was not effective to prevent tectonic corneal graft in cases with severe stromal thinning and impending or recent perforation.

Relevância:

100.00% 100.00%

Publicador:

Resumo:

PURPOSE: To model the possible impact of using average-power intraocular lenses (IOLs) and evaluate the postoperative refractive error in patients having cataract surgery in rural China.SETTING: Rural Guangdong, China.METHODS: Patients having cataract surgery by local surgeons were examined and visual function was assessed 10 to 14 months after surgery. Subjective refraction at near and distance was performed bilaterally by an ophthalmologist. Patients had a target refraction of -0.50 diopter (D) based on ocular biometry.RESULTS: Of the 313 eligible patients, 242 (77%) could be contacted and 176 (74% of contacted patients, 56% overall) were examined. Examined patients had a mean age of 69.4 +/- 10.5 years. Of the 211 operated eyes, 73.2% were within +/-1.0 D of the target refraction after surgery. The best presenting distance vision was in patients within +/-1.0 D of plano and the best presenting near vision, in those with mild myopia (<-1.0 D to > or =2.0 D) (P= .005). However, patients with hyperopia (>+1.0 D) reported significantly better adjusted visual function than those with emmetropia or myopia (<-1.0 D). When the predicted use of an average-power IOL (median +21.5 D) was modeled, predicted visual acuity was significantly reduced (P= .001); however, predicted visual function was not significantly altered (P>.3).CONCLUSIONS: Accurate selection of postoperative refractive error was achieved by local surgeons in this rural area. Based on visual function results, aiming for mild postoperative myopia may not be suitable in this setting. Implanting average-power IOLs significantly reduced postoperative presenting vision, but not visual function.

Relevância:

100.00% 100.00%

Publicador:

Resumo:

PURPOSE: To evaluate the agreement between optical low-coherence reflectometry (OLCR) and anterior segment optical coherence tomography (AS-OCT) for biometry of the anterior segment. SETTING: State Key Laboratory of Ophthalmology, Zhongshan Ophthalmic Center, Sun Yat-sen University, Guangzhou, China. DESIGN: Evaluation of diagnostic technology. METHODS: A series of OLCR (Lenstar LS 900) and AS-OCT measurements of the anterior segment were taken for consecutive subjects aged 35 years and older in a population-based study. The differences and correlations between the 2 methods of ocular biometry were assessed. Agreement was calculated as the 95% limits of agreement (LoA). RESULTS: The mean age of the 776 subjects was 55.2 years ± 12.0 (SD); 54.6% were women. The mean central corneal thickness (CCT) was smaller with OLCR than with AS-OCT (537.84 ± 31.46 μm versus 559.39 ± 32.02 μm) as was anterior chamber depth (ACD) (2.60 ± 0.37 mm versus 2.72 ± 0.37 mm) and anterior chamber width (ACW) (11.76 ± 0.47 mm versus 12.04 ± 0.55 mm) (all P<.001). The 95% LoA between the 2 instruments were -44.80 to 1.71 μm for CCT, -0.17 to -0.06 mm for ACD, and -1.28 to 0.72 mm for ACW. CONCLUSION: Optical low-coherence reflectometry and AS-OCT yielded potentially interchangeable ACD measurements, while the CCT and ACW measurements acquired by the 2 devices showed clinically significant differences.

Relevância:

100.00% 100.00%

Publicador:

Resumo:

PURPOSE: To study, for the first time, the effect of wearing ready-made glasses and glasses with power determined by self-refraction on children's quality of life. METHODS: This is a randomized, double-masked non-inferiority trial. Children in grades 7 and 8 (age 12-15 years) in nine Chinese secondary schools, with presenting visual acuity (VA) ≤6/12 improved with refraction to ≥6/7.5 bilaterally, refractive error ≤-1.0 D and <2.0 D of anisometropia and astigmatism bilaterally, were randomized to receive ready-made spectacles (RM) or identical-appearing spectacles with power determined by: subjective cycloplegic retinoscopy by a university optometrist (U), a rural refractionist (R) or non-cycloplegic self-refraction (SR). Main study outcome was global score on the National Eye Institute Refractive Error Quality of Life-42 (NEI-RQL-42) after 2 months of wearing study glasses, comparing other groups with the U group, adjusting for baseline score. RESULTS: Only one child (0.18%) was excluded for anisometropia or astigmatism. A total of 426 eligible subjects (mean age 14.2 years, 84.5% without glasses at baseline) were allocated to U [103 (24.2%)], RM [113 (26.5%)], R [108 (25.4%)] and SR [102 (23.9%)] groups, respectively. Baseline and endline score data were available for 398 (93.4%) of subjects. In multiple regression models adjusting for baseline score, older age (p = 0.003) and baseline spectacle wear (p = 0.016), but not study group assignment, were significantly associated with lower final score. CONCLUSION: Quality of life wearing ready-mades or glasses based on self-refraction did not differ from that with cycloplegic refraction by an experienced optometrist in this non-inferiority trial.

Relevância:

100.00% 100.00%

Publicador:

Resumo:

Cataract is the leading cause of visual impairment worldwide. In the UK, some 30% of the population over 65 years of age have visually impairing cataract. Importantly, 88% of those with treatable visual impairment from cataract are not in contact with any ocular healthcare service, representing a major potential healthcare need [1]. In the USA, it has been estimated that 17.2% of the population (approximately 20.5 million) over 40 years of age have cataract in either eye and by 2020, this number is expected to rise to 30.1 million. Currently, cataract is responsible for 60% of Medicare costs associated with vision [2]. Furthermore, as the populations of industrialized countries such as the UK and the USA continue to age, the costs associated with treatment of cataract can only be expected to increase. Consequently, the development of the intraocular lens to replace the cataractous lens and the advances in intraocular lens design and implantation represent a major development in cataract treatment. However, despite such advances, cataract surgery is not without complications, such as postoperative infectious endophthalmitis, a rare but potentially devastating condition, and posterior capsular opacification, a less serious but much more common problem. This review will examine the epidemiology of cataracts, the polymeric construction of intraocular lenses implanted during cataract surgery and the complications of postoperative infectious endophthalmitis and posterior capsular opacification with regard to therapeutic interventions and prophylactic strategies. Advances in biomaterial design and function will be discussed as novel approaches to prevent such postoperative complications.

Relevância:

100.00% 100.00%

Publicador:

Resumo:

To estimate the prevalence of refractive error in adults across Europe. Refractive data (mean spherical equivalent) collected between 1990 and 2013 from fifteen population-based cohort and cross-sectional studies of the European Eye Epidemiology (E3) Consortium were combined in a random effects meta-analysis stratified by 5-year age intervals and gender. Participants were excluded if they were identified as having had cataract surgery, retinal detachment, refractive surgery or other factors that might influence refraction. Estimates of refractive error prevalence were obtained including the following classifications: myopia ≤−0.75 diopters (D), high myopia ≤−6D, hyperopia ≥1D and astigmatism ≥1D. Meta-analysis of refractive error was performed for 61,946 individuals from fifteen studies with median age ranging from 44 to 81 and minimal ethnic variation (98 % European ancestry). The age-standardised prevalences (using the 2010 European Standard Population, limited to those ≥25 and <90 years old) were: myopia 30.6 % [95 % confidence interval (CI) 30.4–30.9], high myopia 2.7 % (95 % CI 2.69–2.73), hyperopia 25.2 % (95 % CI 25.0–25.4) and astigmatism 23.9 % (95 % CI 23.7–24.1). Age-specific estimates revealed a high prevalence of myopia in younger participants [47.2 % (CI 41.8–52.5) in 25–29 years-olds]. Refractive error affects just over a half of European adults. The greatest burden of refractive error is due to myopia, with high prevalence rates in young adults. Using the 2010 European population estimates, we estimate there are 227.2 million people with myopia across Europe.

Relevância:

100.00% 100.00%

Publicador:

Resumo:

Purpose: To assess the bacterial contamination risk in cataract surgery associated with mechanical compression of the lid margin immediately after sterilization of the ocular surface.

Setting: Department of Cataract, Zhongshan Ophthalmic Center, Sun Yat-sen University, Guangzhou, China.

Design: Prospective randomized controlled double-masked trial.

Methods: Patients with age-related cataract were randomly assigned to 1 of 2 groups. In Group A (153 eyes), the lid margin was compressed and scrubbed for 360 degrees 5 times with a dry sterile cotton-tipped applicator immediately after ocular sterilization and before povidone-iodine irrigation of the conjunctival sac. Group B (153 eyes) had identical sterilization but no lid scrubbing. Samples from the lid margin, liquid in the collecting bag, and aqueous humor were collected for bacterial culture. Primary outcome measures included the rate of positive bacterial culture for the above samples. The species of bacteria isolated were recorded.

Results: Group A and Group B each comprised 153 eyes. The positive rate of lid margin cultures was 54.24%. The positive rate of cultures for liquid in the collecting bag was significantly higher in Group A (23.53%) than in Group B (9.80%) (P=.001).The bacterial species cultured from the collecting bag in Group B were the same as those from the lid margin in Group A. The positive culture rate of aqueous humor in both groups was 0%.

Conclusion: Mechanical compression of the lid margin immediately before and during cataract surgery increased the risk for bacterial contamination of the surgical field, perhaps due to secretions from the lid margin glands.

Financial Disclosure: No author has a financial or proprietary interest in any material or method mentioned.

Relevância:

100.00% 100.00%

Publicador:

Resumo:

To assess the outcomes of cataract surgery performed by novice surgeons during training in a rural programme. Design: Retrospective study. Participants: Three hundred thirty-four patients operated by two trainees under supervision at rural Chinese county hospitals. Methods: Two trainees performed surgeries under supervision. Visual acuity, refraction and examinations were carried out 3 months postoperatively. Main Outcome Measures: Postoperative uncorrected visual acuity, pinhole visual acuity, causes of visual impairment (postoperative uncorrected visual acuity<6/18) Results: Among 518 operated patients, 426 (82.2%) could be contacted and 334 (64.4% of operated patients) completed the examinations. The mean age was 74.1±8.8 years and 62.9% were women. Postoperative uncorrected visual acuity was available in 372 eyes. Among them, uncorrected visual acuity was ≥6/18 in 278 eyes (74.7%) and <6/60 in 60 eyes (16.1%), and 323 eyes (86.8%) had pinhole visual acuity≥6/18 and 38 eyes (10.2%) had pinhole visual acuity<6/60. Main causes of visual impairment were uncorrected refractive error (63.9%) and comorbid eye disease (24.5%). Comorbid eye diseases associated with pinhole visual acuity<6/60 (n=23, 6.2%) included glaucoma, other optic nerve atrophy, vitreous haemorrhage and retinal detachment. Conclusions: The findings suggest that hands-on training remains safe and effective even when not implemented in centralized training centres. Further refinement of the training protocol, providing postoperative refractive services and more accurate preoperative intraocular lens calculations, can help optimize outcomes. © 2012 The Authors Clinical and Experimental Ophthalmology © 2012 Royal Australian and New Zealand College of Ophthalmologists.

Relevância:

100.00% 100.00%

Publicador:

Resumo:

PURPOSE:

To explore the potential association between adverse corneal events and the use of topical nonsteroidal antiinflammatory drugs (NSAIDs).

SETTING:

Practice-based reports.

METHODS:

A detailed case-reporting form and request for medical records were sent to all practices reporting cases of corneal or conjunctival pathology in association with the use of topical NSAIDs to the American Society of Cataract and Refractive Surgery. Cases were classified as "mild," "moderate," or "severe" according to predetermined clinical criteria.

RESULTS:

Records of 140 eyes (129 patients) were reviewed; 51 cases (36.4%) were mild, 55 (39.3%) moderate, and 34 (24.3%) severe. An association with a specific topical NSAID was confirmed in 117 cases (81.8%). Most confirmed cases (53.8%) involved generic diclofenac (Falcon). Cases associated with brand diclofenac (Voltaren, CIBA Vision) and ketorolac (Acular, Allergan) were more likely to have ocular comorbidity and to have received significantly higher total doses of NSAIDs. Neither "off-label" use nor use of any specific agent was associated with severe compared to mild or moderate disease. However, patients with more severe adverse events were more likely to have a history of diabetes, previous surgery in the affected eye, and surgery other than cataract. Cases not occurring in the perioperative period had significantly worse outcomes, had significantly more ocular comorbidities, and received nearly 3 times the dose of NSAIDs.

CONCLUSIONS:

While topical NSAIDs as a class may be associated with severe adverse events, such events appeared to require potentiation in the form of high total doses, ocular comorbidities, or both with Acular and Voltaren. Severe adverse events might have been more likely to occur at lower doses and in routine postoperative settings with generic diclofenac.

Relevância:

100.00% 100.00%

Publicador:

Resumo:

PURPOSE:

To report determinants of outcomes and follow-up in a large Mexican pediatric cataract project.

SETTING:

Hospital Luis Sanchez Bulnes, Mexico City, Mexico.

METHODS:

Data were collected prospectively from a pediatric cataract surgery program at the Hospital Luis Sanchez Bulnes, implemented by Helen Keller International. Preoperative data included age, sex, baseline visual acuity, type of cataract, laterality, and presence of conditions such as amblyopia. Surgical data included vitrectomy, capsulotomy, complications, and use of intraocular lenses (IOLs). Postoperative data included final visual acuity, refraction, number of follow-up visits, and program support for follow-up.

RESULTS:

Of 574 eyes of 415 children (mean age 7.1 years +/- 4.7 [SD]), IOLs were placed in 416 (87%). At least 1 follow-up was attended by 408 patients (98.3%) (mean total follow-up 3.5 +/- 1.8 months); 40% of eyes achieved a final visual acuity of 6/18 or better. Children living farther from the hospital had fewer postoperative visits (P = .04), while children receiving program support had more visits (P = .001). Factors predictive of better acuity included receiving an IOL during surgery (P = .04) and provision of postoperative spectacles (P = .001). Predictive of worse acuity were amblyopia (P = .003), postoperative complications (P = .0001), unilateral surgery (P = .0075), and female sex (P = .045).

CONCLUSIONS:

The results underscore the importance of surgical training in reducing complications, early intervention before amblyopia (observed in 40% of patients) can develop, and vigorous treatment if amblyopia is present. The positive impact of program support on follow-up is encouraging, although direct financial support may pose a problem for sustainability. More work is needed to understand reasons for worse outcomes in girls.

Relevância:

100.00% 100.00%

Publicador:

Resumo:

PURPOSE: To describe the distribution of central corneal thickness (CCT), intraocular pressure (IOP), and their determinants and association with glaucoma in Chinese adults.DESIGN: Population-based cross-sectional study.METHODS: Chinese adults aged 50 years and older were identified using cluster random sampling in Liwan District, Guangzhou. CCT (both optical [OCCT] and ultrasound [UCCT]), intraocular pressure (by Tonopen, IOP), refractive error (by autorefractor, RE), radius of corneal curvature (RCC), axial length (AL), and body mass index (BMI) were measured, and history of hypertension and diabetes (DM) was collected by questionnaire. Right eye data were analyzed.RESULTS: The mean values of OCCT, UCCT, and IOP were 512 ± 29.0 μm, 542 ± 31.4 μm, and 15.2 ± 3.1 mm Hg, respectively. In multiple regression models, CCT declined with age (P < .001) and increased with greater RCC (P < .001) and DM (P = .037). IOP was positively associated with greater CCT (P < .001), BMI (P < .001), and hypertension (P < .001). All 25 persons with open-angle glaucoma had IOP <21 mm Hg. CCT did not differ significantly between persons with and without open- or closed-angle glaucoma. Among 65 persons with ocular hypertension (IOP >97.5th percentile), CCT (555 ± 29 μm) was significantly (P = .01) higher than for normal persons.CONCLUSIONS: The distributions of CCT and IOP in this study are similar to that for other Chinese populations, though IOP was lower than for European populations, possibly due to lower BMI and blood pressure. Glaucoma with IOP <21 mm Hg is common in this population. We found no association between glaucoma and CCT, though power (0.3) for this analysis was low.

Relevância:

100.00% 100.00%

Publicador:

Resumo:

OBJECTIVE: To compare outcomes between adjustable spectacles and conventional methods for refraction in young people. DESIGN: Cross sectional study. SETTING: Rural southern China. PARTICIPANTS: 648 young people aged 12-18 (mean 14.9 (SD 0.98)), with uncorrected visual acuity ≤ 6/12 in either eye. INTERVENTIONS: All participants underwent self refraction without cycloplegia (paralysis of near focusing ability with topical eye drops), automated refraction without cycloplegia, and subjective refraction by an ophthalmologist with cycloplegia. MAIN OUTCOME MEASURES: Uncorrected and corrected vision, improvement of vision (lines on a chart), and refractive error. RESULTS: Among the participants, 59% (384) were girls, 44% (288) wore spectacles, and 61% (393/648) had 2.00 dioptres or more of myopia in the right eye. All completed self refraction. The proportion with visual acuity ≥ 6/7.5 in the better eye was 5.2% (95% confidence interval 3.6% to 6.9%) for uncorrected vision, 30.2% (25.7% to 34.8%) for currently worn spectacles, 96.9% (95.5% to 98.3%) for self refraction, 98.4% (97.4% to 99.5%) for automated refraction, and 99.1% (98.3% to 99.9%) for subjective refraction (P = 0.033 for self refraction v automated refraction, P = 0.001 for self refraction v subjective refraction). Improvements over uncorrected vision in the better eye with self refraction and subjective refraction were within one line on the eye chart in 98% of participants. In logistic regression models, failure to achieve maximum recorded visual acuity of 6/7.5 in right eyes with self refraction was associated with greater absolute value of myopia/hyperopia (P<0.001), greater astigmatism (P = 0.001), and not having previously worn spectacles (P = 0.002), but not age or sex. Significant inaccuracies in power (≥ 1.00 dioptre) were less common in right eyes with self refraction than with automated refraction (5% v 11%, P<0.001). CONCLUSIONS: Though visual acuity was slightly worse with self refraction than automated or subjective refraction, acuity was excellent in nearly all these young people with inadequately corrected refractive error at baseline. Inaccurate power was less common with self refraction than automated refraction. Self refraction could decrease the requirement for scarce trained personnel, expensive devices, and cycloplegia in children's vision programmes in rural China.

Relevância:

60.00% 60.00%

Publicador:

Resumo:

Histone acetylation is a fundamental mechanism in the regulation of local chromatin conformation and gene expression. Research has focused on the impact of altered epigenetic environments on the expression of specific genes and their pathways. However, changes in histone acetylation also have a global impact on the cell. In this study we used digital texture analysis to assess global chromatin patterns following treatment with trichostatin A (TSA) and have observed significant alterations in the condensation and distribution of higher-order chromatin, which were associated with altered gene expression profiles in both immortalised normal PNT1A prostate cell line and androgen-dependent prostate cancer cell line LNCaP. Furthermore, the extent of TSA-induced disruption was both cell cycle and cell line dependent. This was illustrated by the identification of sub-populations of prostate cancer cells expressing high levels of H3K9 acetylation in the G2/M phase of the cell cycle that were absent in normal cell populations. In addition, the analysis of enriched populations of G1 cells showed a global decondensation of chromatin exclusively in normal cells.